基础医学与临床 ›› 2023, Vol. 43 ›› Issue (5): 809-816.doi: 10.16352/j.issn.1001-6325.2023.05.0809

• 技术与方法 • 上一篇    下一篇

传统与光谱流式细胞仪对自发荧光样本检测的比较

陈家欢1, 李丹丹2, 孙艳1, 王艺蕾3, 王铎3, 孔杰1, 刘晓玲1*   

  1. 1.中国医学科学院基础医学研究所 北京协和医学院基础学院 医学分子生物学国家重点实验室,北京 100005;
    2.中国医学科学院 北京协和医学院 北京协和医院 检验科,北京 100730;
    3.长治医学院,山西 长治 046000
  • 收稿日期:2023-01-17 修回日期:2023-03-21 出版日期:2023-05-05 发布日期:2023-04-26
  • 通讯作者: *happyeday@163.com
  • 基金资助:
    国家重点实验室专项经费(2060204)

Comparative investigation for auto-fluorescence sample detection by evaluating conventional and spectral flow cytometers performance

CHEN Jiahuan1, LI Dandan2, SUN Yan1, WANG Yilei3, WANG Duo3, KONG Jie1, LIU Xiaoling1*   

  1. 1. State Key Laboratory of Medical Molecular Biology, Institute of Basic Medical Sciences CAMS,School of Basic Medicine PUMC, Beijing 100005;
    2. Department of Laboratory Medicine, Peking Union Medical College Hospital, CAMS & PUMC, Beijing 100730;
    3. Changzhi Medical College, Changzhi 046000, China
  • Received:2023-01-17 Revised:2023-03-21 Online:2023-05-05 Published:2023-04-26
  • Contact: *happyeday@163.com

摘要: 目的 比较传统与光谱流式细胞仪检测自发荧光样本的性能。方法 以annexin V(AV)/碘化丙啶(PI)染色检测细胞凋亡为模型,制备带有自发荧光的单细胞样本,分别用传统与光谱流式细胞仪采集同一样本荧光信号。结果 自发荧光干扰异硫氰酸荧光素(FITC)通道荧光信号,传统流式细胞仪应避免使用FITC通道的荧光素;而光谱流式细胞仪可以清楚分辨FTIC通道的荧光素和自发荧光。同时优化了样本制备方法和annexin V/PI染色方案。结论 光谱流式细胞仪对于自发荧光样本的检测在一定条件下具有相对优良的性能。

关键词: 自发荧光, 流式细胞术, 光谱, 异硫氰酸荧光素(FITC)

Abstract: Objective To compare the performance of conventional and spectral flow cytometers in detecting auto-fluorescence samples. Methods Single cell samples with auto-fluorescence were prepared and stained by annexin V(AV)binding and propidium iodure (PI) to detect apoptosis. The same auto-fluorescence sample was run on conventional and spectral flow cytometers side by side. Results By avoid using FITC in conventional cytometers, the better result was gotten in detecting auto-fluorescence samples. Full spectrum cytometry presented better resolution even there was spectral overlap. It was easy to discriminate the signals derived from auto-fluorescence and FITC. The protocol of annexinV binding and PI staining to analyze apoptosis was optimized. Conclusions Spectral flow cytometry shows relatively excellent performance under certain conditions in detecting of apoptosis derived from auto-fluorescence samples.

Key words: auto-fluorescence, flow cytometry, spectrum, fluorescein isothiocyanate(FITC)

中图分类号: