Studies on test method of the effective deposition for inhalation

Jin Fng

Chinese Pharmaceutical Journal ›› 1999, Vol. 34 ›› Issue (10) : 672-674.

Chinese Pharmaceutical Journal ›› 1999, Vol. 34 ›› Issue (10) : 672-674.
药剂

Studies on test method of the effective deposition for inhalation

  • Jin Fang
Author information +
History +

Abstract

OBJECTIVE:To study on method of inhalation simulation test in vitro.The effective dose of inhalation was deter mined.METHOD:The imitated twin-stage liq wuid impinger was used to dtermine the effective dose of inhalation.There was 7ml receiving liquid in stageⅠ,and 30 ml in stageⅡ (simulated action position).The aspirated speed was 60L·min-1.The test results were compared with that reported.RESULTS:There sults were corresponded with the foreign reports.The effective dose of inhalation was related to the dosage form,active ingredient properties and the receiving liquid kings.CONCLUSION:The imitated twin-stage liquid impinger could be used to control the quality of inhalation product.

Key words

inhalation / the effective dose / twin-stage liquid impinger

Cite this article

Download Citations
Jin Fng. Studies on test method of the effective deposition for inhalation[J]. Chinese Pharmaceutical Journal, 1999, 34(10): 672-674

References

1 Hallworth GW.The twin impinger:a simple device for assessing the delivery of drtlgs from meterde dose pressurized aerosol inhalers.J Pharmacol,1987,39:966. 2 Martin GP,Bell AE,Marriott C.An in vitro method for assessing particle deposition from metered pressurised aerosols and dry powder inhalers.Int J Pharm,1988,44(1):57.