在线柱萃取HPLC和柱前衍生HPLC测定丙戊酸血药浓度的比较研究

李威, 彭敏, 姚敦武

中国药学杂志 ›› 2018, Vol. 53 ›› Issue (20) : 1785-1788.

PDF(1345 KB)
PDF(1345 KB)
中国药学杂志 ›› 2018, Vol. 53 ›› Issue (20) : 1785-1788. DOI: 10.11669/cpj.2018.20.016
论著

在线柱萃取HPLC和柱前衍生HPLC测定丙戊酸血药浓度的比较研究

  • 李威1, 彭敏2*, 姚敦武1
作者信息 +

Comparison of On-line Column Extraction HPLC and Pre-column Derivatization HPLC Method for Determination of Valproic Acid in Plasma

  • LI Wei1, PENG Min2*, YAO Dun-wu1
Author information +
文章历史 +

摘要

目的 比较在线柱萃取高效液相色谱法(HPLC)与柱前衍生HPLC测定人血中丙戊酸(VPA)浓度的差异及相关性,为临床VPA血药浓度监测提供参考依据。方法 分别用2种方法对120份患者血样的VPA浓度进行检测,采用统计学方法对其检测结果的差异进行评价。结果 以柱前衍生HPLC测定结果(X)与在线柱萃取HPLC测定结果(Y)所作线性回归方程如下:Y=0.804 2X-2.611 3(r2=0.973 9),2种方法测定VPA血药浓度结果的差异有统计学意义(P<0.05)。结论 2种方法均可用于临床VPA的血药浓度监测,但对测得结果的差异应予以密切关注;相比柱前衍生HPLC,在线柱萃取HPLC操作简单且不易受其代谢物干扰,更适于临床VPA的血药浓度监测。

Abstract

OBJECTIVE To investigate the differences and correlation between the methods of on-line column extraction high performance liquid chromatography(HPLC) and pre-column derivatization HPLC for the determination of valproic acid(VPA) in plasma and provide reference for the clinical monitoring of VPA concentration. METHODS One hundred and twenty samples were detected by the two methods respectively, and the differences of the test results were evaluated by statistical methods. RESULTS The linear regression equation between the results determined by the pre-column derivatization HPLC method(X) and the on-line column HPLC method(Y) was as follows:Y=0.804 2X-2.611 3(r2=0.973 9). There was significant difference between the results determined by the two methods(P<0.05). CONCLUSION Both methods can be used to monitor the concentration of VPA in clinical practice. However, it is necessary to pay attention to the difference between the two methods. And the on-line column HPLC method is more suitable for monitoring the concentration of VPA in clinical practice because it is simple to operate and has less interference from the metabolites.

关键词

丙戊酸 / 在线柱萃取高效液相色谱法 / 柱前衍生高效液相色谱法 / 治疗药物监测

Key words

valproic acid / on-line column extraction HPLC / pre-column derivatization HPLC / therapeutic drug monitoring

引用本文

导出引用
李威, 彭敏, 姚敦武. 在线柱萃取HPLC和柱前衍生HPLC测定丙戊酸血药浓度的比较研究[J]. 中国药学杂志, 2018, 53(20): 1785-1788 https://doi.org/10.11669/cpj.2018.20.016
LI Wei, PENG Min, YAO Dun-wu. Comparison of On-line Column Extraction HPLC and Pre-column Derivatization HPLC Method for Determination of Valproic Acid in Plasma[J]. Chinese Pharmaceutical Journal, 2018, 53(20): 1785-1788 https://doi.org/10.11669/cpj.2018.20.016
中图分类号: R965   

参考文献

[1] JOHANNESSEN S I, LANDMARK C J. Value of therapeutic drug monitoring in epilepsy. Exp Rev Neurother, 2008, 8(6):929-939.
[2] GLAUSER T, BEN-MENACHEM E, BOURGEOIS B, et al. Updated ILAE evidence review of antiepileptic drug efficacy and effectiveness as initial monotherapy for epileptic seizures and syndromes. Epilepsia, 2013, 54(3):551-563.
[3] XU S S, ZHANG T, LIU M, et al. The linfluence of therapeutic drug monitoring compliance on efficiency and safety about valproic acid in epileptic children. Chin Pharm J(中国药学杂志), 2015, 50(20):1829-1832.
[4] ORY D, VAN DEN BRANDE J, DE GROOT T, et al. Retention of fluoride on reversed phase HPLC columns. J Pharm Biomed Anal, 2015, 111(10):209-214.
[5] ZHANG J F, ZHANG Z Q, DONG W C, et al. A new derivatization method to enhance sensitivity for the determination of low levels of valproic acid in human plasma. J Chromatographic Sci, 2014, 52(10):1173-1180.
[6] LUAN W L, ZHU L T, YANG L, et al. Advances in the determination of valproic acid in human plasma by pre-column derivatization HPLC . J China Pharm(中国药房), 2016, 27(11):1577-1579.
[7] LI X, WANG F, XU B, et al. Determination of the free and total concentrations of vancomycin by two-dimensional liquid chromatography and its application in elderly patients. J Chromatogr B, 2014,969:181-189.
[8] BOBALY B, TOTH E, DRAHOS L, et al. Influence of acid-induced conformational variability on protein separation in reversed phase high performance liquid chromatography. J Chromatogr A, 2014,1(1325):155-162.
[9] SATHEESHKUMAR N,NAVEENKUMAR G. A stability-indicating reversed-phase high-performance liquid chromatography method for Ambrisentan:an endothelin receptor antagonist. J Chromatogr Sci, 2014, 52(8):894-898.
[10] HU X F, ZHANG Y J, WEI L H. Analysis the influencing factors on plasma concentration of valproic acid sodium. Chin Hosp Pharm J(中国医院药学杂志), 2017, 37(13):1246-1255.
[11] CHEM Z J, WANG X D, WANG H S, et al. Simultaneous determination of valproic acid and 2-propyl-4-pentenoic acid for the prediction of clinical adverse effects in Chinese patients with epilepsy. Seizure, 2012, 21(2):110-117.
[12] GAO S H, MIAO H J, XIA T, et al. LC-MS/MS Method for simultaneous determination of valproic acid and major metabolites in human plasma. J Chromatogr B, 2011, 879(21):1939-1944.
[13] WANG C, KONG A Y, ZHOU B T, et al. Simultaneous determination of valproic acid and its hepatotoxic metabolite by gas chromatography-mass spectrometry. Chin Pharm J(中国药学杂志), 2015, 50(14):1222-1227.
PDF(1345 KB)